Model 1800 amplifier – A-M Systems
The A-M Systems Model 1800 amplifier is a 2-channel AC differential amplifier with preamplifier (headstage), designed for extracellular recording and stimulation applications with high-impedance microelectrodes.
Description of the A-M Systems Model 1800 amplifier
Ideal for recording single-cell action potentials (spikes), EEG, EMG, EKG, ERG and evoked potentials in research and teaching laboratories.
Features of the A-M Systems Model 1800 amplifier
- 2 independent channels, including headstages
- Fixed gain ×100, ×1,000, ×10,000
- Low noise: ~1.0 µV peak-to-peak (10 Hz-10 kHz)
- Common mode rejection > 80 dB (up to 110 dB depending on source)
- Effective notch filter (> 25 dB at 50 Hz / > 30 dB at 60 Hz)
- High input impedance > 100,000 MΩ
- Low bias current ± 3 pA typical, max ± 20 pA
- High-pass filters: 0.1, 1, 10, 100, 300 Hz
- Low-pass filters: 500, 1 k, 5 k, 10 k, 20 k Hz
- Modes: Record, Stimulate, Impedance check (in situ)
- Current measurement during stimulation via internal 1 kΩ resistor
- Sampling heads included with 1.5 m rigid cables
- Power supply: 110 V/60 Hz or 220 V/50 Hz
- Standard 3-year warranty
Technical characteristics of the A-M Systems Model 1800 amplifier
| Caractéristique | Valeur |
|---|---|
| Channels | 2 differential amplifiers with headstages |
| Gain | ×100, ×1 000, ×10 000 |
| Noise | 1.0 µV p-p typical (10 Hz-10 kHz) |
| Common mode rejection | > 80 dB (up to 110 dB at 60 Hz) |
| Notch filter | > 25 dB at 50 Hz, > 30 dB at 60 Hz |
| Input impedance | > 100,000 MΩ |
| Polarisation current | ± 3 pA typical, ± 20 pA max |
| High-pass filters | 0.1 / 1 / 10 / 100 / 300 Hz |
| Low-pass filters | 500 Hz / 1 k / 5 k / 10 k / 20 k Hz |
| Mode selection | Record / Stimulate / Impedance Check |
| Internal resistance (Stim/Impede.) | 1 kΩ (current/stimulation) |
| Headstage | 0.535″ Ø × 4″, rod 4.7 mm, cable 1.5 m |
| Power supply | 110 V / 60 Hz or 220 V / 50 Hz |
| Dimensions (W×H×D) | 43.2 × 12.1 × 28.6 cm |
| Consumption | < 5 W |
| Accessories included | 2 headstages, cables, rack mount, manual |
| Guarantee | 3 years |
Applications
- Extracellular recordings of cellular spikes, multi-units
- EEG, EMG, EKG, ERG and evoked potentials
- Verification of electrode impedance in situ
- Research and teaching in neurophysiology, nerves and muscles