Integrated patching system (amplifier + data acquisition + software) dPatch
The dPatch® system is available in single or dual preamplifier versions, offering the possibility of recording simultaneously at impressed potential and impressed current. Each pre-amplifier operates independently and samples data at 5 MHz with 18-bit resolution at 1 MHz. The outputs are filtered at 13 settings, from 100 Hz to 1 MHz. Its low-noise design allows stable recordings at room temperature without the need to cool the pre-amplifier. The system includes eight analogue input channels, four analogue outputs and sixteen digital outputs, with a high sampling rate, and allows acquisition to be triggered via an internal clock or external TTL. SutterPatch software allows the addition and monitoring of over 600 experimental metadata, stimulus application, instrument control and real-time data analysis. Configuration is quick and connections simplified for easy set-up.
Features of the dPatch integrated patch system (amplifier + data acquisition + software)
- Patch-clamp amplifier integrated into the A/D converter
- Versions available: dPatch (one preamplifier) and dPatch-2 (two independent preamplifiers)
- Pre-amplifiers with three sensitivity ranges for whole cells or single channels
- Low RMS noise (0.1-10 kHz): dPatch < 2.3 pA; dPatch-2 < 0.22 pA
- Automatic compensation for electrode and cytoplasmic membrane capacitance
- Recording in imposed potential and imposed current modes
- High bandwidth for fast signal detection
- Compensation of series resistors
- Pre-amplifier sampling at 5 MHz, 18-bit resolution at 1 MHz
- 13-setting filtered outputs (100 Hz to 1 MHz)
- Multiplexerless acquisition: 8 analogue input channels (±10 V, 200 kHz), 4 analogue outputs (±10 V, 250 kHz), 16 digital outputs (TTL, 250 kHz)
- Acquisition triggered by on-board clock or external trigger (TTL)
- Low noise at ambient temperature without cooling
- SutterPatch software for control, stimulus application, data analysis and management of over 600 metadata
- Dimensions: dPatch 48.2 × 28 × 9 cm; pre-amplifier 19.5 × 9 × 3 cm; head 10 × 3.5 × 1.9 cm
- Weight: 6.8 kg
Applications
- Recordings in single channel configuration
- Recordings of cells in slice and in primary culture
- Cell line studies (in culture or dispersed)
- Optogenetics