MICRO MÉCANIQUE offers a comprehensive range of microscopy solutions, cameras, lenses, lighting, motorised stages, incubation chambers and accessories. We support you from selection to installation, including training and maintenance, to optimise your observations and protocols, from teaching laboratories to advanced research.

Electrophysiology & Neuroscience

For over 35 years, MICRO MÉCANIQUE SAS has been supporting researchers and laboratories in neuroscience, electrophysiology and optogenetics. We offer a comprehensive range of equipment (amplifiers, micromanipulators, microinjection systems, etc.) and comprehensive support: advice, installation, training and maintenance throughout France.

Model 2100 insulated stimulator

The Model 2100 is an isolated pulse stimulator with integrated optical and transformer isolation, capable of delivering single or two-phase pulses in constant current or voltage mode up to ±100 V or ±10 mA. It operates in free-running, manual or external trigger (TTL) mode, with status LED indicators (TRIG, EVENT, ERROR). It offers exceptional time accuracy thanks to a 10 MHz clock, for timing from 1 µs to over 15 minutes.

Features of the Model 2100 Stimulator

  • Free-running, manual or external TTL trigger operation
  • LED indicators TRIG (trigger received), EVENT (pulse in progress), ERROR (delivery error)
  • Pulse train settings via lever-wheel: delay, train width, pulse duration, interpulse period
  • Duration ranges from 1 µs to 999 s (~15 min) with accuracy < 0.02 % and jitter ≤ ±250 ns
  • Output modes: current or voltage (switchable), with independent amplitude for pulse and baseline
  • Single-phase (±10 mA / ±100 V) or two-phase (±5 mA / ±50 V) pulses as selected
  • User-friendly thumbwheel interface and no need for external isolator

Technical specifications of the Model 2100 Stimulator

Caractéristique Valeur
Timing range 1 µs to > 999 s (~15 minutes)
Time accuracy < 0.02 % + jitter ≤ ±250 ns
Output Current or voltage mode (selectable)
Single-phase amplitude ±10 mA or ±100 V
Two-phase amplitude ±5 mA or ±50 V
Output impedance Voltage mode < 60 Ω ; Current mode >1 MΩ
Amplitude adjustment 10-turn potentiometers + switch range
Trigger modes Free-run, manual, TTL external trigger
LED indicators TRIG (trigger), EVENT (pulse), ERROR (error delivery)
Sync outputs TRIG OUT and BURST signal for external synchronisation
Insulation Transformer + integrated opto-coupling
Power supply External transformer (no battery)Installation
Assembly Rack mount hardware included
Guarantee 3 years manufacturer’s standard

Applications

  • Neurophysiology stimulation protocols: electrode marker, LTP/LTD, kindling, orthodromic/antidromic
  • Experiments requiring high temporal precision (fast trains, very short durations)
  • Protocols requiring electrical isolation to avoid artefacts or interference
  • Stand-alone or controlled via computer or external footswitch

Documentation

Model 2100 brochure